Explain the process you employ to evaluate
the condition of a device.

IMS performs a comprehensive evaluation of every device received at our repair facility. The following criteria are part of the evaluation for incoming rigid endoscopes:

Items tested at the Optical Bench


  • Field of view
  • Angle of view
  • Angle of view to bore (mechanical axis)
  • Illumination view angle
  • Overlapping of illumination and field of view
Optical bench
 

Items tested using a Diopter Telescope


  • Magnification
  • Depth of view
Diopter Telescope
 

Items tested using 1956 U.S. Air Force Resolution Targets
or customized IMS Targets


  • Resolution
Resolution target
 

Items tested using a Light Meter / Ulbricht Sphere


  • Light transmission efficiency of the fibers and optical system
Ulbricht Sphere
 

Items tested using a microscope


  • Observation of mechanical defects to include optics - visual inspection
Microscope
 

More: Disassembly and Repair Process »